| SPECIFICATIONS | 
| MODEL | 7305 | 7314 | 7315 | 7316 | 
| GROUND BOND | 
| Output AC Current | 3 - 30A | 1 - 40A | 1 - 30A | 1 - 60A | 
| Output AC Voltage | 6V, Maximum | 8V, Maximum | 6V, Maximum | 12V, Maximum | 
| Output Frequency | 50Hz/60Hz ±100ppm, User Selection | 
| SETTINGS | 
| Max-LimitRange
 | Resistance | 0-510mΩ
 | 0-120mΩ
 | 0-600mΩ
 | 0-200mΩ
 | 0-150mΩ
 | 0-510mΩ
 | 0-120mΩ
 | 0-600mΩ
 | 0-300mΩ
 | 0-150mΩ
 | 
| AC Current | 3.0-10.0A
 | 10.1-30.0A
 | 1.00-10.00A
 | 10.01-30.00A
 | 30.01-40.00A
 | 1.00-10.00A
 | 10.1-30.00A
 | 1.00-15.00A
 | 15.01-30.00A
 | 30.01-60.00A
 | 
| Resolution | 1mΩ | 
| Accuracy | Same as Resistance Measurement Accuracy | 
| Min-LimitRange
 | Resistance | - | 0-600mΩ
 | 0-200mΩ
 | 0-150mΩ
 | 0-510mΩ
 | 0-120mΩ
 | 0-600mΩ
 | 0-300mΩ
 | 0-150mΩ
 | 
| AC Current | - | 1.00-10.00A
 | 10.01-30.00A
 | 30.01-40.00A
 | 1.00-10.00A
 | 10.1-30.00A
 | 1.00-15.00A
 | 15.01-30.00A
 | 30.01-60.00A
 | 
| Resolution | 1mΩ | 
| Accuracy | Same as Resistance Measurement Accuracy | 
| Lead Resistance Offset | Range | 0 - 100 mΩ, Auto/manual | 
| Resolution | 1mΩ | 
| Accuracy | ±(2% of setting +2 counts) | 
| Dwell Time | Range | 0, 0.5 - 999.9 sec (0 = continuous) | 
| Resolution | 0.1 sec | 
| Accuracy | ±(0.1% + 0.05sec) | 
| MEASUREMENT | 
| AC Current | Range | 3.0 - 30.0A | 1.00 - 40.00A | 1.00 - 30.00A | 1.00 - 60.00A | 
| Resolution | 0.1A | 0.01A | 0.01A | 0.01A | 
| Accuracy | ±(2%of reading +1count) | ±(3%of reading +3counts) | ±(3%of reading +3counts) | ±(3%of reading +3counts) | 
| Resistance | Range | 0-510mΩ
 | 0-120mΩ
 | 0-600mΩ
 | 0-600Ω
 | 0-200mΩ
 | 0-150mΩ
 | 0-510mΩ
 | 0-510mΩ
 | 0-120mΩ
 | 0-600mΩ
 | 0-600mΩ
 | 0-300mΩ
 | 0-150mΩ
 | 
| 3.0-10.0A
 | 10.1-30.0A
 | 1.00-5.99 A | 6.00-10.00 A
 | 10.01-30.00 A
 | 30.01-40.00 A
 | 1.00-2.99 A
 | 3.00-10.00 A
 | 10.01-30.00 A
 | 1.00-5.99 A
 | 6.00-15.00 A
 | 15.01-30.00 A
 | 30.01-60.00 A
 | 
| Resolution | 1 mΩ | 
| Accuracy | ±(2%of reading +1count) | ±3% of reading +3 counts) | ±(2% of reading +2 counts) | ±(3% of reading +3counts) | ±(2% of reading +2 counts) | ±3% of reading +3 counts) | ±(2% of reading +2 counts) | 
| GENERAL | 
| Input Voltage AC | 115/230Vac±15%, 50/60Hz ± 5%, Fuse 6.3A for 7305/7315 model, 10A for 7314 model, 15A for 7316 model  | 
| PLC Remote Control | Input: Test , Reset,Withstand Processing ,(Memory1.2.3, Interlock for 7314/7315/7316 model only) | 
| Output: Pass , Fail , Processing , Reset Out , Start Out | 
| Memory | 5 memories | 10 memories, 3 steps/memory | 
| Display | 16×2 LCD with back light | 128×64 Graphic LCD  | 
| Key Lock | To prevent unauthorized alteration of the test Parameters | 
| Calibration | Build-in software and external calibrated meters | 
| Verification | - | Build-in software driven verification menu to test fault detection circuits | 
| Alarm Volume Setting | - | Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. | 
| Environment | 0-40℃, 20-80%RH | 
| Dimension | 280(W)×89(H)×370(D)mm
 | 215(W)×89(H)×430(D)mm
 | 215(W)×89(H)×370(D)mm
 | 430(W)×130(H)×400(D)mm
 | 
| Net Weight | 10Kg | 14Kg | 8.7Kg | 20Kg | 
| STANDARD ACCESSORIES | 
| Power Cord(10A) | ×1  | ×1 (15A) | 
| Fuses | ×2 (Including a spare contained in the fuse holder) | ×1(On Board) | 
| Interlock Disable Key(1505) | - | ×1  | ×1  | ×1  | 
| Ground Bond Test Lead 30A ,1.5m(1103) | - | - | ×1  | - | 
| Ground Bond Return Lead 30A ,1.5m(1104) | - | - | ×1  | - | 
| Ground Bond Return Lead 30A ,1.6m(1122) | ×1  | - | - | - | 
| Ground Bond Test Lead 30A,1.6m(1131) | ×1  | - | - | - | 
| Ground Bond Test Lead 40A,1.6m(1137) | - | ×1  | - | - | 
| Ground Bond Return Lead 40A,1.6m(1138) | - | ×1  | - | - | 
| Ground Bond Test Lead 60A,1.65m (1160) | - | - | - | ×1  | 
| Ground Bond Return Lead 60A,1.65m(1161) | - | - | - | ×1 |