| SPECIFICATIONS | 
| MODEL | 7130 | 7132 | 7140 | 7142 | 
| AC WITHSTAND VOLTAGE | 
| Output Rating  | 5KVAC/20mA | 
| Range | Resolution | Accuracy | 
| Output Voltage, KVAC | 0-5.00 | 0.01 | ±(2% of setting + 5V) | 
| Output Frequency | 50Hz/60Hz ±100ppm, User Selection | 
| Output Waveform | Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 | 
| Output Regulation | ± (1% of output + 5V), From no load to full load  | 
| SETTINGS | 
| Max-Limit AC Current, mA | 0 - 20.00 | 0.01 | ±(2% of setting + 2counts) | 
| Min-Limit AC Current, mA | 0 - 9.999 | 0.001 | ±(2% of setting + 2counts) | 
| Ramp Up Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | 
| Ramp Down Time, second | 0 - 999.9 | 0.1 | 
| Dwell Time, second | 0, 0.3 - 999.9(0=continuous)
 | 0.1 | 
| Arc Detection | 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)  | 
| DC WITHSTAND VOLTAGE (7140,7142 ONLY) | 
| Output Rating  | 6KVDC/7500μA | 
| Output Voltage, KVDC | 0 - 6.00 | 0.01 | ±(2% of setting + 5V) | 
| Output Ripple | <5%(6KV/7500μA at Resistive Load) | 
| SETTINGS | 
| Max-Limit DC Current, μA | 0 - 7500 | 1 | ±(2% of setting + 2counts) | 
| Min-Limit DC Current, μA | 0 - 999.9 | 0.1 | ±(2% of setting + 2counts) | 
| Ramp Up Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | 
| Ramp Down Time, second | 0, 1.0 - 999.9 | 0.1 | 
| Dwell Time, second | 0, 0.4 - 999.9(0=continuous)
 | 0.1 | 
| Arc Detection | 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity)  | 
| Discharge Time | < 200 msec | 
| INSULATION RESISTANCE (7132,7142 ONLY) | 
| Output Rating  | 1KVDC/9999MΩ | 
| Output Voltage, DCV | 30 - 1000 | 10 | ±(2% of setting + 5V) | 
| SETTINGS | 
| Max-Limit Resistance, MΩ | 0, 1 - 9999 (0=OFF) | 1 | ±(2% of setting+2counts)at 500-1000VDC, 1 -1000 MΩ
 | 
| Min-Limit Resistance, MΩ | 1 - 9999 | 1 | ±(5% of setting+2counts)at 500-1000VDC, 1000 -9999 MΩ
 | 
| ±(8% of setting+2counts)at 30-500VDC, 1 -1000 MΩ
 | 
| Ramp Up Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | 
| Ramp Down Time, second | 0, 1.0 - 999.9 | 0.1 | 
| Delay Time, second | 0, 0.5 - 999.9(0=continuous)
 | 0.1 | 
| MEASUREMENT | 
| AC/DC Voltage, KV | 0 - 6.00 | 0.01 | ±(1.5% of reading + 1count) | 
| DC Voltage, V (IR only) | 30 - 1000 | 1 | ±(1.5% of reading + 5V) | 
| AC Current, mA | 0 - 3.500 | 0.001 | ±(2% of reading + 2counts) | 
| 3.00 - 20.00 | 0.01 | 
| DC Current, μA | 0 - 350.0 | 0.1 | ±(2% of reading + 2counts) | 
| DC Current, mA | 0.300 - 3.500 | 0.001 | 
| 3.00 - 7.50 | 0.01 | 
| Resistance, , MΩ | 1 - 9999(AutoRange)
 | 0.0010.01
 0.1
 1
 | ±(2% of setting+2counts)at 500-1000VDC,1 -1000 MΩ
 ±(5% of setting+2counts)at 500-1000VDC,
 1000 -9999 MΩ
 ±(8% of setting+2counts)at 30-500VDC,
 1 -1000 MΩ
 | 
| GENERAL | 
| Input Voltage AC | 115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A | 
| PLC Remote Control | Input : Test , Reset , Memory1.2.3 , Interlock | 
| Output : Pass , Fail , Processing , Reset-Out  | 
| Interface(Option) | RS485 Interface | 
| Memory | 10 memories, 3 steps/memory | 
| Graphic Display | 128×64 Graphic LCD  | 
| Safety | Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS | 
| Key Lock | To prevent unauthorized alteration of the test Parameters | 
| Calibration | Build-in software and external calibrated meters | 
| Verification | Build-in software driven verification menu to test fault detection circuits | 
| Alarm Volume Setting | Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. | 
| Environment | 0-40℃, 20-80%RH | 
| Dimension/Net Weight | 215mm(W)×89mm(H)×370mm(D)/ 10Kg | 
| STANDARD ACCESSORIES | 
| Power Cord(10A) | ×1  | 
| Fuses | ×2 (Including a spare contained in the fuse holder) | 
| Interlock Disable Key(1505) | ×1  | 
| Hipot Test Lead ,1.5m(1101) | ×1  | 
| Hipot Return Lead ,1.8m(1102) | ×1  | 
| *product specifications are subject to change without notice. |