| SPECIFICATIONS | 
| MODEL | 7730 | 
| AC WITHSTAND VOLTAGE | 
|   | Range | Resolution | Accuracy | 
| Output Voltage, ACV | 0-5000 | 1 | ±(2% of setting+ 5V) | 
| Output Frequency | 50Hz/60Hz ±100ppm, User Selection | 
| Output Waveform | Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 | 
| Output Regulation | ± (1% of output + 5V), From no load to full load  | 
| SETTINGS | 
| Maximum Limit current, mA | 0.000 - 9.999 (0=OFF) | 0.001 | ± (2% of setting + 2 counts) | 
| 10.00-30.00 | 0.01 | 
| Minimum Limit current, mA | 0.000 - 9.999 (0=OFF) | 0.001 | ± (2% of setting + 2 counts) | 
| 10.00-30.00 | 0.01 | 
| Ramp-Up Timer, second | 0.1 - 999.9 | 0.1 | ± (0.1% + 0.05 sec) | 
| Ramp-Down Timer, second | 0.0 - 999.9 | 
| Dwell Timer, second | 0, 0.4 - 999.9(0=continuous)
 | 
| Arc Detection | 1 - 9 ranges (9 is the most sensitivity)  | 
| DC WITHSTAND VOLTAGE | 
| Output Voltage, DCV | 0-5000 | 1 | ±(2% of setting+ 5V) | 
| Output Ripple | <4 %(5KV/10mA at Resistive Load) | 
| SETTINGS | 
| Maximum Limit, current, μA | 0 - 999.9 | 0.1 | ± (2% of setting + 2 counts) | 
| 1000 - 10000  | 1 | 
| Minimum Limit, current, μA | 0.0 - 999.9 | 0.1 | ± (2% of setting + 2 counts) | 
|   | 1000 - 10000  | 1 | 
| Ramp-Up Timer, second | 0.1 - 999.9 | 0.1 | ± (0.1% + 0.05 sec) | 
| Ramp-Down Timer, second | 0, 1.0 - 999.9 (0=OFF) | 
| Dwell Timer, second | 0, 0.4 - 999.9(0=continuous)
 | 
| Ramp-HI, current  | >12 mA peak maximum, ON/OFF User Selection | 
| Charge-LO, current | 0.0 - 350.0 μA DC or Auto Set | 
| Discharge Time | < 200 msec | 
| Maximum Capacitive LoadDC Mode | 1 μF < 1KV 0.08 μF < 4KV | 
| 0.75 μF < 2KV 0.04 μF < 5KV | 
| 0.5 μF < 3KV  | 
| Arc Detection | 1 - 9 ranges (9 is the most sensitivity)  | 
| INSULATION RESISTANCE | 
| Output Voltage, VDC | 50 - 1000 | 1 | ± (2% of setting + 2 counts) | 
| Short Circuit Current | Maximum >12mA peak | 
| SETTINGS | 
| Maximum and Minimun Limit Resistance, MΩ
 | 0.05 - 99.99(Max – Limit: 0 = OFF) | 0.01 | Same as Resistance MEASUREMENT Accuracy | 
| 100.0-999.9 | 0.1 | 
| 1000-50000 | 1 | ± (0.1% + 0.05 sec) | 
| Ramp-Up Timer, second | 0.1 - 999.9 | 0.1 | 
| Ramp-Down Timer, second | 0, 1.0 - 999.9 (0=OFF) | 
| Delay Timer, second | 0, 1.0 - 999.9(0=continuous)
 | 0.1 | 
| Charge-LO, current, μA | 0.000 - 3.500 or Auto Set | 
| MEASUREMENT | 
|   | Range | Resolution | Accuracy | 
| Voltage, KV (AC/DC) | 0.00-5.00 | 0.01 | ± (1.5% of reading +2counts) | 
| Voltage, Vdc (IR only) | 0-1000 | 1 | ± (1.5% of reading + 2counts) | 
| AC Current, mA | 0.000-3.500 | 0.001 | ± (2% of reading + 2counts) | 
| 3.00-30.00 | 0.01 | 
| DC Current, μA | 0.0 - 350.0 | 0.1 | ± (2% of reading + 2 counts) | 
| DC Current, mA | 0.300 - 3.500 | 0.001 | 
| 3.00-9.99 | 0.01 | 
| Resistance, MΩ | 50-499V | 500-1000V |   | 50 – 499V0.05–999.9, ±(7% of reading+2 counts)
 | 
| 0.050-1.999 | 0.050-9.999 | 0.001 | 500 – 1000V | 
| 2.00-19.99 | 10.00-99.99 | 0.01 | 0.10–999.9, ±(2% of reading+2 counts) | 
| 20.0-199.9 | 100.0-999.9 | 0.1 | 1000–9999, ±(5% of reading+2 counts) | 
| 200-50000 | 1000-50000 | 1 | 10000–50000, ±(15%of reading+2 counts) | 
| GENERAL | 
| Input Voltage | 115/230Vac±15% Auto Selection, 50/60Hz ± 5%, Fuse 4A Slow-Blo 250VAC | 
| PLC Remote Control | Input: Test, Reset, Interlock, Recall Memory 1 through 10 | 
| Output: Pass, Fail, Test-in-Process, Reset Out | 
| Remote Output Relays | 125VAC@1AAC, 30VDC@ 0.5ADC | 
| Memory | 50 memories, 30 steps/memory | 
| Graphic Display | 240 x 64 dot resolution Monographic LCD /Contrast 9 Levels 1-9 | 
| Safety | Built-in Smart GFI circuit,GFI trip current 450μA max, HV shut down speed: <1mS | 
| Interface | Standard RS232, Optional PrinterPort with Date and Time Stamp or GPIB (IEEE-488.2). | 
| Security | Programmable password lockout capability to avoid unauthorized access to test set-up program. | 
| Alarm Volume Setting | Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. | 
| Calibration | Adjustments are made through the fornt panel. Built-in Verification program. | 
| Automatic Calibration alert function to signal operator when calibration is due. | 
| Environment | 0-40℃, 20-80%RH | 
| Dimensions/Net Weight | (W × H × D) 2U (430 × 89 × 400 mm)/13Kg. | 
| Option (Opt.735) | DC Continuity Test Function (0.1A/ 0-2KΩ) at Rear Panel | 
| STANDARD ACCESSORIES | 
| Power Cord(10A) | ×1  | 
| Fuses | ×2 (Including a spare contained in the fuse holder) | 
| Interlock Disable Key(1505) | ×1  | 
| Hipot Test Lead ,1.5m(1101) | ×1  | 
| Hipot Return Lead ,1.8m(1102) | ×1  | 
| RS232 Link Cable 1.5m(1130) | ×1  | 
| MATRIX SCANNER MODULE (Opt.729) | 
| High Voltage Rating | 5KVAC/5KVDC | 
| Number of HV Channel | 8 | 
| MATRIX SCANNER MODULE (Opt.730) | 
| High Voltage Rating | 5KVAC/5KVDC | 
| Number of HV Channel | 4 | 
| DC CONTINUITY TEST FUNCTION (Opt.735) | 
| MEASUREMENT | 
|   | Range | Resolution | Accuracy | 
| Output Current | DC 0.1A ± 0.01A, Max. | 
| Resistance Display | 0.00 – 19.99  | 0.01 | ± (1 % of reading + 5 counts) | 
| 20.0 – 199.9  | 0.1 | ± (1 % of reading + 2 counts) | 
| 200 – 2000  | 1 | ± (1 % of reading + 2 counts)  | 
| SETTINGS | 
| Max-Limit, Ω | 0.00 – 19.99  | 0.01 | ± (1 % of setting + 5 counts) | 
| 20.0 – 199.9  | 0.1 | ± (1 % of setting + 2 counts) | 
| 200 – 2000  | 1 | ± (1 % of setting + 2 counts)  | 
| Min-Limit, Ω | 0.00 – 19.99 | 0.01 | ± (1 % of setting + 5 counts) | 
| 20.0 – 199.9  | 0.1 | ± (1 % of setting + 2 counts) | 
| 200 – 2000  | 1 | ± (1 % of setting + 2 counts)  | 
| Dwell Timer, second | 0.0, 0.3 – 999.9 (0 = continuous)
 | 0.1 | ± (0.1% + 0.05 sec) | 
| Resistance Offset, Ω | 0.00 – 2.00 | 0.01 | ± (1 % of reading + 5 counts) |