53/4-Digit Display (480,000 Counts)
Simultaneous Display of 3 Measurements,
e.g. DC + AC + Statistics
Up to 200 Measurements per Second
DC Basic Accuracy 0,015%
12 Measurements Functions: DCV, DCI, True RMS, ACV and ACI, Frequency, 2- and 4-Wire Resistance, Capacitance, Continuity, Diode Test, Temperature, Power
Crisp color TFT display for excellent readability R Resolution: 1µV, 100nA, 1mΩ, 1pF, 1Hz, 0,1°C R True RMS Measurement AC, AD + DC
Mathematic Functions: Limit Testing, Minimum/Maximum, Average, Offset, DC Power, dB, dBm
Temperature Measurements with Platinum Sensors (PT100/PT500/PT1000)
Data Logging in .CSV-Format to Internal Memory or USB-Stick
Interfaces: USB-TMC and Ethernet (LXI in Preparation), optional IEEE-488 (GPIB)
SCPI commands widely compatible with Agilent 34410A
DC Specifications DC Accuracy in ± (% of reading + % of range): |
Function | Range1) | Test Current Voltage drop | Input Impedance | 1 Year 23 °C ± 5 K | Temp. Coefficient 0-18 °C, 28-55 °C |
DC Voltage | 400,000 mV | | 10 MΩ/>10 GΩ | 0,015+0,002 | 0,0010+0,0005 |
| 4,00000 V | | 10 MΩ/>10 GΩ | 0,015+0,002 | 0,0008+0,0003 |
| 40,0000 V | | 10 MΩ | 0,020+0,002 | 0,0010+0,0005 |
| 400,000 V | | 10 MΩ | 0,020+0,002 | 0,0015+0,0005 |
| 1000,00 V | | 10 MΩ | 0,025+0,002 | 0,0015+0,0005 |
Resistance | 400,000 Ω | 1 mA | | 0,050+0,005 | 0,0020+0,0005 |
(2/4-wire) 2) | 4,00000 kΩ | 1 mA | | 0,015+0,002 | 0,0020+0,0002 |
| 40,0000 kΩ | 100 µA | | 0,015+0,002 | 0,0020+0,0002 |
| 400,000 kΩ | 10 µA | | 0,030+0,003 | 0,0020+0,0002 |
| 4,00000 MΩ | 1 µA | | 0,060+0,005 | 0,0020+0,0002 |
| 40,0000 MΩ | 100 nA | | 0,250+0,003 | 0,0080+0,0005 |
| 250,000 MΩ | 460nA||10MΩ (parallel) | | 2,000+0,010 | 0,200+0,0005 |
DC Current 4) | 20,0000 mA | <0,30 V | | 0,05+0,010 | 0,008+0,0010 |
| 200,000 mA | <0,27 V | | 0,05+0,010 | 0,008+0,0010 |
| 2,00000 A | <0,2 V | | 0,25+0,070 | 0,012+0,0015 |
| 10,0000 A 3) | <0,60 V | | 0,25+0,070 | 0,010+0,0010 |
Continuity | 4000 Ω | 1 mA | | 0,05+0,010 | 0,005+0,0005 |
Diode Test | 5 V | 1 mA | | 0,05+0,010 | 0,005+0,0005 |
AC Voltage 2) | 400,000 mV | 10 Hz–20 Hz | 3,0+0,05 | 0,01+0,01 |
| 4,00000 V | 20 Hz–45 Hz | 1,5+0,05 | 0,01+0,01 |
| 40,0000 V | 45 Hz–20 kHz | 0,3+0,05 | 0,01+0,01 |
| 400,000 V | 20 kHz–50 kHz | 1,0+0,05 | 0,02+0,01 |
| 750,00 V 6) | 50 kHz–100 kHz | 3,0+0,05 | 0,05+0,01 |
AC Current 5) | 20,0000 mA | 20 Hz–40 Hz | 1,5+0,05 | 0,01+0,01 |
| 200,000 mA | 40 Hz–1 kHz | 0,5+0,05 | 0,01+0,01 |
| 2,00000 A | 1 kHz–5 kHz | 1,5+0,05 | 0,01+0,01 |
| 10,0000 A 4) | 5 kHz–10 kHz 3) | 2,5+0,05 | 0,01+0,01 |
Frequency Counter Specifications Frequency Accuracy in ± (% of reading): |
Function | Range1) | Frequency | 1 Year 23 °C ± 5 K | Temp. Coefficient 0-18 °C, 28-55 °C |
AC Voltage 2) | all ranges | 5 Hz–700 kHz | 0,01 | 0,005 |
AC Current 2) | 20 mA, 200 mA | 5 Hz–10 kHz | 0,01 | 0,005 |
| 2 A, 10 A | 5 Hz–5 kHz | 0,01 | 0,005 |
Frequency Counter Resolution |
Setting | Measurement time | Display range | Resolution |
Slow | 1 s | 999,999 kHz | 1 Hz |
Medium | 100 ms | 999,99 kHz | 10 Hz |
Fast | 10 ms | 999,9 kHz | 100 Hz |