Voltage Specifications1, 2,

SourceMeasure3
Range4ResolutionAccuracy5 23 °C ± 5 °C,
1 Year ±(% setting + volts)
Noise (RMS)<10 HzResolutionInput ResistanceAccuracy 23 °C ± 5 °C,
1 Year ±(% reading + volts)
200.0000 mV5 µV0.015% + 200 µV2 µV100 nV> 10 GΩ0.012% + 200 µV
2.000000 V50 µV0.020% + 300 µV10 µV1 µV> 10 GΩ0.012% + 300 µV
20.00000 V500 µV0.015% + 2.4 mV100 µV10 µV> 10 GΩ0.015% + 1 mV
200.0000 V5 mV0.015% + 24 mV1 mV100 µV> 10 GΩ0.015% + 10 mV
1000.000 V50 mV0.02% + 100 mV20 mV10 mV> 10 GΩ0.015% + 50 mV

Temperature Coefficient 20 V and 1000 V ranges: ± (0.15 × accuracy specification)/°C, 0 °C to 18 °C and 28 °C to 50 °C
200 mV and 2 V ranges: ±(0.30 x accuracy specification)/°C, 0 °C to 18 °C and 28 °C to 50 °C


Current Specifications1, 2

SourceMeasure3
Range4ResolutionAccuracy523 °C ± 5 °C,
1 Year ±(% setting + amps)
Noise (RMS)<10 HzResolutionVoltage BurdenAccuracy23 °C ± 5 °C,
1 Year ±(% reading + amps)
10.00000 nA6500 fA0.100% + 200 pA500 fA10 fA<100 µV0.10% + 250 pA
100.0000 nA65 pA0.060% + 250 pA500 fA100 fA<100 µV0.060% + 300 pA
1.000000 μA50 pA0.025% + 400 pA5 pA1 pA< 100 µV0.025% + 300 pA
10.00000 μA500 pA0.025% + 1.5 nA40 pA10 pA<100 µV0.025% + 700 pA
100.0000 μA5 nA0.020% + 15 nA400 pA100 pA<100 µV0.02% + 6 nA
1.000000 mA50 nA0.020% + 150 nA5 nA1 nA<100 µV0.02% + 60 nA
10.00000 mA500 nA0.020% + 1.5 μA40 nA10 nA<100 µV0.02% + 600 nA
100.0000 mA5 μA0.025% + 15 μA100 nA100 nA<100 µV0.025% + 6 μA
1.000000 A50 μA0.067% + 900 μA10 μA1 μA<100 µV0.03% + 500 μA

Temperature Coefficient ± (0.15 × accuracy specification)/°C, 0 °C to 18 °C and 28 °C to 50 °C


Notes

1. Speed = 1 PLC.

2. All specifications are guaranteed with output ON.

3. Accuracies apply to 2-wire and 4-wire modes when properly zeroed. For the 200 mV and 1 A ranges, the voltage burden may exceed the specification in 2-wire mode.

4. Maximum display and programming ranges are 5% overrange for voltage, except for the 1000 V range, which is 10% overrange (1100 V), and 5% overrange for current (for example,1.05 A on the 1 A range).

5. For sink mode, 1 µA to 100 mA range, accuracy is ±(0.5% + offset × 3). For 1 A range, accuracy ± (0.15% x accuracy specification))/°C on 20 V and 1000 V ranges ±(1.5% + offset × 3) .

6. Rear-panel triaxial connections only


Resistance Measurement Accuracy (Local Or Remote Sense)7, 8, 9
RangeDefault Resolution10Default Test CurrentNormal Accuracy 23 °C ± 5 °C,
1 Year ± (% reading + ohms)
Enhanced Accuracy11 23 °C ± 5 °C,
1 Year ± (% reading + ohms)
< 2.000000 Ω121 µΩUser-definedSource IACC+ Meas VACCMeas IACC + Meas VACC
20.00000 Ω10 µΩ100 mA0.098% + 0.003 Ω0.073% + 0.001 Ω
200.0000 Ω100 µΩ10 mA0.077% + 0.03 Ω0.053% + 0.01 Ω
2.000000 kΩ1 mΩ1 mA0.066% + 0.3 Ω0.045% + 0.1 Ω
20.00000 kΩ10 mΩ100 µA0.063% + 3 Ω0.043% + 1 Ω
200.0000 kΩ100 mΩ10 µA0.065% + 30 Ω0.046% + 10 Ω
2.000000 MΩ1 Ω1 µA0.11% + 300 Ω0.049% + 100 Ω
20.00000 MΩ10 Ω1 µA0.11% + 1 kΩ0.052% + 500 Ω
200.0000 MΩ 13100 Ω100 nA0.655% + 10 kΩ0.349% + 5 kΩ
>200.0000 MΩ12, 13-User-definedSource IACC+ Meas VACCMeas IACC+ Meas VACC
Temperature Coefficient± (0.15 × accuracy specification)/°C 0 °C to 18 °C and 28 °C to 50 °C
Source Current, Measure Resistance ModeTotal uncertainty = I source accuracy + V measure accuracy (4-wire remote sense)
Source Voltage, Measure Resistance ModeTotal uncertainty = V source accuracy + I measure accuracy (4-wire remote sense)
Guard Output Impedance≥ 300 Ω typical

Notes

7. Speed = 1 PLC.

8. All specifications are guaranteed with output ON.

9. Accuracies apply to 2-wire and 4-wire modes when properly zeroed.

10. Measure resolution 6.5 digits.

11. Source readback enabled; offset compensation on.

12. Source current, measure resistance or source voltage, measure resistance only.

13. Rear-panel triaxial connections only.


Supplemental Specifications
Overrange105% of range for 200 mV to 200 V, source and measure ranges; 110% of range for 1000 V, source and measure ranges
Regulation
    VoltageLine: 0.01% of range
Load: 0.01% of range + 100 µV
  
    CurrentLine: 0.01% of range
Load: 0.01% of range + 100 pA
Source Limits
    Voltage Source Current LimitBipolar current limit set with a single value
Minimum value is 10% of range
  
    Current Source Voltage LimitBipolar voltage limit set with a single value
Minimum value is 10% of range
Voltage Limit/Current Limit AccuracyAdd 0.3% of range and ±0.02% of reading to base specification
Overshoot
    Voltage Source<0.1% typical
Step size = Full scale, resistive load, 20 V range, 10 mA current limit
  
    Current Source<0.1% typical
Step size = Full scale, resistive load of 10 kΩ, 1 mA range, 20 V voltage limit
Range Change OvershootOvershoot into a fully resistive 100 kΩ load, 10 Hz to 20 MHz bandwidth, adjacent ranges: 250 mV typical
Output Settling TimeTime required to reach 0.1% of final value after command is processed and output slew: 20 V range,100 mA current limit: < 200 µs typical
Maximum Slew Rate140.2 V/µs, 200 V range, 100 mA limit into a 2 kΩ load (typical)
0.5 V/µs, 1000 V range, 10 mA limit into a 100 kΩ load (typical)
Overvoltage ProtectionUser-selectable values, 10% tolerance; factory default = none
Voltage Source Noise10 Hz to 20 MHz (RMS): 4 mV typical into a resistive load
Common Mode Voltage250 V DC
Common Mode Isolation> 1 GΩ, < 1000 pF
Noise Rejection (typical)NPLCNMRRCMRR0.01-60 dB0.1-60 dB160 dB100 dB*

* Except lowest two current ranges ~90dB.

Load Impedance
    Normal Mode20 nF typical
  
    High-Capacitance ModeStable into 1 μF typical (specification only valid for ranges ≥100 μA)
Maximum Voltage Drop Between Force and Sense Terminals5 V
Maximum Sense Lead Resistance1 MΩ for rated accuracy
Sense Input Impedance> 10 GΩ
Guard Offset Voltage< 300 µV typical

Notes

14. High capacitance mode off.


System Measurement Speeds15

Reading rates (readings per second) typical for 60 Hz (50 Hz), script (TSP) programmed

NPLCTrigger OriginMeasure to MemoryMeasure to GPIBMeasure to USBMeasure to LANSource Measure Sweep to MemorySource measure sweep to GPIBSource Measure Sweep to USBSource Measure Sweep to LAN
0.01Internal3150 (2800)2760 (2570)2825 (2570)2740 (2530)1710 (1620)1620 (1540)1630 (1540)1620 (1540)
0.01External2170 (2050)2120 (2003)2170 (2010)2100 (1990)1670 (1590)1580 (1500)1590 (1510)1580 (1510)
0.10Interna540 (460)530 (450)530 (450)530 (450)470 (410)460 (400)470 (400)470 (400)
0.10External500 (430)490 (420)500 (425)480 (420)450 (400)460 (390)460 (390)410 (350)
1.00Internal59 (4958 (49)59 (49)59 (49)58 (48)58 (48)58 (48)57 (48)
1.00External58 (48)57 (48)58 (48)58 (48)57 (48)57 (48)57 (48)55 (48)

Reading rates (readings per second) typical for 60 Hz (50 Hz), SCPI programmed16

NPLCTrigger OriginMeasure to MemoryMeasure to GPIBMeasure to USBMeasure to LANSource Measure Sweep to memorySource Measure Sweep to GPIBSource Measure Sweep to USBSource Measure Sweep to LAN
0.01Internal3040 (2800)3000 (2760)3000 (2760)3010 (2710)1710 (1630)1610 (1544)1440 (1380)1690 (1590)
0.01External2320 (2165)2290 (2140)2340 (2150)2290 (2130)1680 (1590)1560 (1525)1410 (1360)1660 (1560)
0.10Internal540 (460)540 (450)540 (460)540 (450)540 (410)470 (410)450 (390)470 (410)
0.10External510 (440)510 (430)510 (440)510 (430)470 (400)470 (400)450 (390)470 (400)
1.00Internal59 (49)59 (49)59 (49)59 (49)58 (48)58 (48)57 (48)58 (48)
1.00External58 (49)58 (49)58 (49)58 (49)58 (49)58 (48)57 (47)58 (48)

Notes

15. Reading rates applicable for voltage or current measurements, autozero off, autorange off, filter off, binary reading format, and source readback off.

16. SCPI programming mode.


General Characteristics (default mode unless specified)

Factory Default Standard Power-UpSCPI Mode
Source Output ModesFixed DC Level, Memory/Configuration List (mixed function), Stair (linear and log).
Memory Buffer>5,000,000 readings with selected measured values and timestamp.
Real-Time ClockLithium battery backup (3 yr. + battery life).
Remote InterfacesGPIBIEEE-488.1 compliant. Supports IEEE-488.2 common commands and status model topology.USB Device (rear panel, type B)2.0 Full Speed USBTMC.USB Host (front panel, type A)USB 2.0, support for flash drives, FAT32.EthernetRJ-45 (10/100BT).
Digital I/O Interface:Lines6 Input/Output user defined for digital I/O or triggering.Connector9-pin female D.Input Signal Levels0.7 V (maximum logic low), 3.7 V (minimum logic high).Input Voltage Limits-0.25 V (Abs. minimum), +5.25 V (Abs. maximum).Maximum Source Current+2.0 mA @ >2.7 V (per pin).Maximum Sink Current-50 mA at 0.7 V (per pin, solid-state fuse protected)5V Power Supply PinLimited to 500 mA @ >4 V (solid-state fuse protected).HandlerUser definable Start of Test, End of Test, 4 category bits.
ProgrammabilitySCPI or TSP command sets.
TSP ModeEmbedded Test Script Processor (TSP) accessible from any host interface.
IP ConfigurationStatic or DHCP.
Expansion InterfaceThe TSP-Link expansion interface allows TSP enabled instruments to trigger and communicate with each other.
LXI Compliance1.5 LXI Device Specification 2016.
Display5 inch capacitive touch, color TFT WVGA (800×480) with LED backlight.
Input Signal ConnectionsFront: Banana. Rear: High-voltage triaxia.
InterlockActive High Input.
CoolingForced air, variable speed.
Over Temperature ProtectionInternally sensed temperature overload puts unit in standby mode.
Power Supply100 V to 240 V RMS, 50–60 Hz (automatically detected at power up).
VA Rating190 volt-amps max.
AltitudeMaximum 2000 meters above sea level.
EMCConforms to European Union EMC Directive.
SafetyNRTL listed to UL61010-1 and UL61010-2-30. Conforms with European Union Low Voltage Directive.
VibrationMIL-PRF-28800F Class 3 Random.
Warm-Up1 hour to rated accuracies.
DimensionsWith bumpers and handle: 106 mm high × 255 mm wide × 425 mm deep (4.18 in × 10.05 in × 16.75 in).
Without bumpers and handle: 88 mm high × 213 mm wide × 403 mm deep (3.46 in × 8.39 in × 15.87 in.)
WeightWith bumpers and handle: 4.54 kg (10 lb).
Without bumpers and handle: 4.08 kg (9.0 lb).
EnvironmentOperating: 0 °C to 50 °C, 70% relative humidity up to 35 °C; derate 3% relative humidity/°C, 35 °C to 50 °C.
Storage: -25 °C to 65 °C.
Pollution Category: 2